Chlorine Dioxide

Electrochemical sensor for measurements of chlorine dioxide.
The application areas of this sensor extend to almost all water qualities. It is resistant to chemicals and detergents thanks to a special membrane system. The chlorine dioxide sensor is also resistant to chlorine. Ozone is measured with a 25 times higher sensitivity than chlorine dioxide. The measuring cell can be used in the pH range from pH >1 up to the limit of stability of chlorine dioxide in alkaline solutions. Precipitation, such as lime, can block the membrane! The Sensor has a plug and is not delivered with a fixed cable.
Measurement principle
Amperometry
Measuring range
0...2 mg/L, 0...20 mg/L
Measurement accuracy
Measurement range 2 mg/L: at 0.4 mg/L; 1.6 mg/L < 1%
Measurement range 20 mg/L: at 1.5 mg/L < 0.1%
Temperature compensation
Automatic through integrated temperature sensor Pt100
Response time
T90: ~ 1 min
Housing material
PVC-U, Stainless steel 1.4571
Dimensions (L x Ø)
205 mm x 25 mm
Weight
0.12 kg
Interface
RS-485, Modbus RTU
Power supply
12...24 VDC (± 10 %)
Connector
8-pin M12-plug

Power consumption
0.6 W
Maintenance effort
≤ 0.5 h/month (typical)
Calibration/maintenance interval
Membrane cap change: every 24 months;
Electrolyte change: every 3 - 6 months
System compatibility
Modbus RTU
Warranty
1 Year (EU & US: 2 years) on electronics; All wearing parts are not included in the warranty

Max. pressure
1 bar
~ 14.5 psig
Protection type
IP68
NEMA 6P
Probentemperatur
+5...+50 °C
~ +41 °F... +122 °F
Umgebungstemperatur
+5...+55 °C
~ +41 °F... +131°F
Lagertemperatur
+5...+35 °C
~ +41 °F... +95 °F